Z-0355-2027: Yeastone Broth, 11ML, 10/BOX, used in antibiotic susceptibility testing of mycological clinical specimens

Recall record Z-0355-2027 for Remel, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0355-2027
Company in FDA record
Remel, Inc
FDA classification date
Nov 3, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Affected lot did not meet specifications. Product may report false susceptibility to antifungal agents, specifically within the Azole family.

Source facts

Action
The firm began notifying affected customers on October 19, 2026. Customers were instructed to notify any personnel who needed to be aware of the issue. Customers should review previously reported susceptibility results generated using the affected lot and assess whether retesting is appropriate. Customers should destroy any remaining inventory of the affected lot. Do not use it for any clinical laboratory testing. Customers should complete and return the acknowledgement form accompanying the letter. Firm will replace product or issue credit. If you have any questions, please contact our Technical Services Department at 800-255-6730 (US) or email microbiology.ts.us@thermofisher.com. You should complete the accompanying Acknowledgment Form upon receipt and return to MBD.vigilance@thermofisher.com
Initiated
2026-10-19
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of MA and CA. The countries of India, Saudi Arabia, UK.
Recall number
Z-0355-2027
Recall status
Open, Classified
CDRH record ID
223399
FDA classified
2026-11-03
Recall event number
100139

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗