Multiple defects: 1) beds exhibit reverse motions when the Fowler, gatch or foot section is activated 2) a battery charging defect which may result in a depletion of battery power 3) the touch screen on the footboard freezing, resulting in controls to powered, pressure-relief mattresses being difficult to access 4) beds equipped with zoom drive systems may have an overdrive problem which could re
Source facts
Action
Consignees were notified via an Urgent Medical Device Correction letter dated 11/2/07, and instructed to remove the beds from service if they display any unintended motions and informed that one of the firm's field service representatives will visit the facility to make corrections to all beds under recall.
Initiated
2007-11-02
Root cause
Process control
Terminated
2010-10-27
Distribution
Worldwide-USA and Canada
Recall number
Z-0356-2008
Recall status
Terminated
CDRH record ID
65476
FDA classified
2007-12-08
Recall event number
45445