Open original FDA record ↗Recall record
- FDA identifier
- Z-0356-2014
- Company in FDA record
- Restoration Robotics
- FDA classification date
- Nov 18, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Reports of discoloration (residue) on the pins of the skin tensioner have been received by several customers. Discoloration due to insufficient cleaning of stainless steel pins during passivation process. The residue has been identified as a compound of iron and chlorine.
Source facts
- Action
- Restoration Robotics sent an Urgent Recall letters dated October 13, 2013 to all distributors via first class mail. Direct accounts were notified by phone as well as fax. Customers were informed of the affected product, problem and actions to be taken. Replacement products are being distributed to each site and will be delivered to each customer. For questions call 408-883-6760
- Initiated
- 2013-10-13
- Root cause
- Process control
- Terminated
- 2013-12-10
- Distribution
- Worldwide Distribution - USA Nationwide in the countries of: Korea, Japan, Singapore, Thailand, United Kingdom, Canada, and Taiwan.
- Recall number
- Z-0356-2014
- Recall status
- Terminated
- CDRH record ID
- 122606
- FDA classified
- 2013-11-18
- Recall event number
- 66558