Z-0356-2022: BV Pulsera with Software Release 2.3-A mobile, diagnostic X-ray image acquisition and viewing system. Mod

Recall record Z-0356-2022 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0356-2022
Company in FDA record
Philips North America Llc
FDA record creation date
Dec 9, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Instructions for Use do not specify the maximum surface temperature of the X-ray tank (assembly housing the X-ray tube, beam filter & cooling oil) and the image intensifier/detector of the system as required an result in burns

Source facts

Action
Philips issued an Urgent Medical Device Correction letter in the USA on 11/12/21 via Certified mailing United States Postal Service. Recalling outside the US is done through the local Philips organizations in the respective countries Letter states reason for recall, health risk and action to take: Pay attention to the Heat Indication Level Icons shown in the Viewing Station to know the temperature of the X-Ray Tank. o Follow the instructions provided in Annex I of this letter for a safe use of the system. o Circulate this notice to all users of this device so that they are aware of the issue. o Place this Field safety Notice with the documentation of the of your system o Transfer this Field Safety notice within your organization or to any organization, if the affected devices have been transferred. o Complete the enclosed Customer Reply Form and send it back to confirm that users have reviewed and understood this Urgent Field Safety Notice. 5. Actions planned by Philips IGT Systems to correct the problem Philips is informing customers through this Urgent Field Safety Notice and providing the Addendum to the Instructions For Use of the BV Endura Release 2.3, BV Pulsera Release 2.3 and Veradius Unity Release 2.1 systems.
Initiated
2021-11-12
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Algeria, American Samoa, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Cameroon, Chile, China, Colombia, Cook Islands, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Faroe Islands, Finland, France, French Guiana, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guyana, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kuwait, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Malaysia, Maldives Mali, Martinique, Mauritius, Mexico, Moldova, Monaco, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of Panama, Papua New Guinea, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia, Zimbabwe.
Recall number
Z-0356-2022
Recall status
Open, Classified
CDRH record ID
190586
FDA record created
2021-12-09
Recall event number
89129

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗