Skip to content

Public FDA record

Z-0356-2024: stryker Sage PrimaFit External Urine Management System for the Female Anatomy

Recall record Z-0356-2024 for Sage Products Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0356-2024
Company in FDA record
Sage Products Inc
FDA classification date
Nov 21, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A small percentage of devices in three lots of product may contain a tape with a natural rubber latex adhesive which is not declared on the label

Source facts

Action
Stryker issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 10/23/2023 via USPS Certified Mail. The notice explained the issue and the potential risk to user and requested the following actions: 1. Locate the product listed above and if tape is detected on the product once the product is opened, do not use on patients that have a latex allergy. Isolate and discard any impacted product with tape to avoid accidental use by a patient with a latex sensitivity. 2. Distribute this notice to all hospital staff that need to be aware of the problem with this device. 3. Maintain awareness of this communication internally by placing this notice at the point where the product is stored prior to use. 4. Inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients appropriately. Additional action for distributors only: 1. Locate the product listed above in your inventory and attach a provided label to the side to each case of the impacted product to ensure that the End User will be made aware of this issue.
Initiated
2023-10-23
Root cause
Process change control
Distribution
US and Canada
Recall number
Z-0356-2024
Recall status
Open, Classified
CDRH record ID
203729
FDA classified
2023-11-21
Recall event number
93229

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗