Open original FDA record ↗Recall record
- FDA identifier
- Z-0356-2027
- Company in FDA record
- CORIN MEDICAL, LTD.
- FDA classification date
- Nov 3, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.
Source facts
- Action
- Corin notified consignees on about 10/02/2026 via emailed "URGENT MEDICAL DEVICE REMOVAL" letter. Consignees were instructed to identify the products present in inventory and quarantine them, return the devices to Corin (replacement parts will be provided), provide the notification letter to all involved parties, and complete and return the acknowledgement form. Additionally, Corin requested notification if affected units have been further distributed as well as any affected customers so that they may be notified.
- Initiated
- 2026-10-02
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX, UT and the countries of Belgium, France, Germany, Hungary.
- Recall number
- Z-0356-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222992
- FDA classified
- 2026-11-03
- Recall event number
- 99959