Skip to content

Public FDA record

Z-0357-2016: IOL Master 5.5: Software versions 7.5 and 7.7; Ophthalmic: used to obtain ocular measurements and perfor

Recall record Z-0357-2016 for Carl Zeiss Meditec AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0357-2016
Company in FDA record
Carl Zeiss Meditec AG
FDA classification date
Dec 1, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

IOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.

Source facts

Action
Urgent Medical Device Correction letters will be mailed on Friday, November 13, 2015 as the IOL SW update material is now available. Letter informs customers of the issue and provide instructions for identifying the software version and for installing the software update with illustrations. All customers are requested to complete and return the confirmation form to Stericycle as soon as the upgrade is complete or it has been determined that your unit is not running the affected version of the software.
Initiated
2015-10-27
Root cause
Software design
Terminated
2016-06-06
Distribution
Nationwide Distribution.
Recall number
Z-0357-2016
Recall status
Terminated
CDRH record ID
141530
FDA classified
2015-12-01
Recall event number
72564

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗