IOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.
Source facts
Action
Urgent Medical Device Correction letters will be mailed on Friday, November 13, 2015 as the IOL SW update material is now available. Letter informs customers of the issue and provide instructions for identifying the software version and for installing the software update with illustrations.
All customers are requested to complete and return the confirmation form to Stericycle as soon as the upgrade is complete or it has been determined that your unit is not running the affected version of the software.
Initiated
2015-10-27
Root cause
Software design
Terminated
2016-06-06
Distribution
Nationwide Distribution.
Recall number
Z-0358-2016
Recall status
Terminated
CDRH record ID
141531
FDA classified
2015-12-01
Recall event number
72564
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.