Z-0358-2024: GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330

Recall record Z-0358-2024 for Instrumentation Laboratory. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0358-2024
Company in FDA record
Instrumentation Laboratory
FDA classification date
Nov 21, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 9/29/23 was sent to customers. Mandatory Customer Actions Please take the following immediate actions: " Check your inventory for the GEM Premier 3000 Non-iQM PAK (cartridge) lot numbers listed in Appendix A. " Discontinue use and destroy any GEM Premier 3000 Non-iQM PAK (cartridge) lot numbers listed in Appendix A. " Document the destruction on the Mandatory Tracking Response Record and return the completed and signed form to Werfen Regulatory at the fax number or e-mail address listed below. " Contact Werfen Customer Service at 1-800-955-9525 to order replacement PAKs (cartridges). " Share this information with your laboratory staff and follow your internal procedures. " Forward this notification to all affected locations within your facility. " Retain a copy of this notification for your records. Mandatory Response Tracking Record " Return the completed and signed Mandatory Response Tracking Form to: " Fax No.: 1-781-861-4207 Or " E-mail: ra-usa.na@werfen.com Contact Information for Questions " For technical questions, contact the Technical Support Center at 1-800-678-0710. " For replacement cartridges, contact Customer Service at 1-800-955-9525.
Initiated
2023-09-29
Root cause
Under Investigation by firm
Distribution
US States: MA, MN, NY. India.
Recall number
Z-0358-2024
Recall status
Open, Classified
CDRH record ID
203891
FDA classified
2023-11-21
Recall event number
93313

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗