Open original FDA record ↗Recall record
- FDA identifier
- Z-0359-2011
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- Nov 15, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The power supply board in GE Healthcare Dash 3000, 4000 and 5000 patient monitors may fail due to insufficient solder of a component. Insufficient solder on the power supply board could result in excess EMC emissions, smoking of components and eventual failure of the power supply, tripping of the fuse and immediate shutdown resulting in a loss of monitoring that could delay recognition of a clinic
Source facts
- Action
- GE Healthcare issued an "Urgent Medical Device Correction" letter dated October 8, 2010, which described the loss of monitoring issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. Customers were instructed to remove the units from use. Local service representatives will contact firms to schedule installation of replacement power supply boards. GE Healthcare can be contacted at 1-800-558-7044, Option 2, Technical Support, about this correction.
- Initiated
- 2010-10-11
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-05-29
- Distribution
- Worldwide Distribution: USA in the states of IL, NV, TX, and VA, and the countries of Australia, Brazil, China, Japan, South Korea, and Taiwan.
- Recall number
- Z-0359-2011
- Recall status
- Terminated
- CDRH record ID
- 94202
- FDA classified
- 2010-11-15
- Recall event number
- 56664