Skip to content

Public FDA record

Z-0360-2014: Salter Labs EtCO2 Sample line 15' (4.6 M) tube, male luer connectors. Rx only; Made in Mexico; Manufactur

Recall record Z-0360-2014 for SALTER LABS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0360-2014
Company in FDA record
SALTER LABS
FDA classification date
Nov 19, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It has been determined that three lots of model number 4515-15-25 CO2 Sample Line 15' tube, male Luer lock connectors may contain a 30' tube instead of the 15' length.

Source facts

Action
Salter Labs called the five customers that received product from the three lots, then follow-up immediately by sending a recall notification letter directly to each customer. Each letter will identify all shipments of subject product that they received. All customers letters were sent certified mail, return receipt requested.
Initiated
2013-11-01
Root cause
Labeling Change Control
Terminated
2013-12-17
Distribution
Distributed in the states of MS, TX FL,, WI, MI, VA, NY, MO, NC, and MS.
Recall number
Z-0360-2014
Recall status
Terminated
CDRH record ID
123119
FDA classified
2013-11-19
Recall event number
66720

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗