Open original FDA record ↗Recall record
- FDA identifier
- Z-0360-2016
- Company in FDA record
- Draeger Medical, Inc.
- FDA classification date
- Dec 12, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
There is a potential problem with the device main power switch of the Perseus A500 Anesthesia Workstation. The device power switch may spontaneously fail during use, causing one of the following to occur: 1. The workstation has a fresh gas delivery failure and a ventilator failure with respective alarms. 2. The workstation shuts down completely with an audible alarm.
Source facts
- Action
- Draeger Medical, Inc. sent Urgent Medical Device Recall Letters, dated November 2015, to consignees on November 10, 2015. The letter identified the affected device and the reason for the recall. The device power switch of the affected workstations must be replace. Customers should refer to the list of affected serial numbers provided. A Drager Service Representative will contact customers to schedule the replacement free of charge, as soon as new switches are available. Until the replacement takes place, the workstation may only be operated under continuous supervision. Increased attention is necessary to notice the potential failure case early. All affected users should be informed. Customers can contact Michael Kelhart at 1-800-543-5047, (press 1 at the prompt, then 2, then 32349), with questions pertaining to the letter. Contact Drager Service Technical Support at 1-800-543-5047 for questions regarding the operation or servicing of the Drager Perseus A500.
- Initiated
- 2015-11-12
- Root cause
- Nonconforming Material/Component
- Terminated
- 2017-04-05
- Distribution
- US Distribution -- To the states of WI, PA, OH, ME, IL, and NY.
- Recall number
- Z-0360-2016
- Recall status
- Terminated
- CDRH record ID
- 141791
- FDA classified
- 2015-12-12
- Recall event number
- 72649