Open original FDA record ↗Recall record
- FDA identifier
- Z-0361-2017
- Company in FDA record
- Hospira Inc.
- FDA record creation date
- Nov 4, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
May go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The infusion and the processor stop, sound a 3 minute audible alarm from the backup buzzer and deactivate the couch screen and the automated loading/unloading of cassettes. Delay or interruption of therapy.
Source facts
- Action
- Hospira sent an Urgent Device Recall Letters, dated February 5, 2013, to all affected customers. Customers were directed to notify their consignees if they further distributed the product and confirm they have done so by return of the reply form. Healthcare professionals were to contact Stericycle. Hospira discontinued support of these devices as of June 30, 2015. Customers with questions were instructed to call 1-800-201-9064. For questions regarding this recall call 224-212-4892.
- Initiated
- 2011-03-16
- Root cause
- Device Design
- Terminated
- 2016-11-28
- Distribution
- Worldwide Distribution - US (nationwide) and Canada
- Recall number
- Z-0361-2017
- Recall status
- Terminated
- CDRH record ID
- 115201
- FDA record created
- 2016-11-04
- Recall event number
- 64005