Z-0361-2019: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) (same product) FlexLab, Accelerator a360

Recall record Z-0361-2019 for Inpeco S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0361-2019
Company in FDA record
Inpeco S.A.
FDA record creation date
Oct 31, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has identified customer sites where there are modules on which the safety pacemaker label is missing. The information regarding the label included in the Operations Manual has been updated to specify that: 1) the risk of a pacemaker malfunction applies to anyone with a pacemaker who may work on these modules; and 2) the minimum safety distance has been increased to 200 mm (7.87") from the transport mechanisms. Any person with a pacemaker must not get closer than this distance from the modules axes.

Source facts

Action
On 8/24/18, the firm, Inpeco, notified distributors of the recall via an "Urgent IVD Medical Device Recall" letter dated August 2018. Distributors were informed of the product issue and instructed to further notify end users. Advise on action to be taken by the user: 1. Please be aware that personnel fitted with a pacemaker must not handle or work on these modules at distances lower than 200 mm even if the warning label is missing. 2. Please review the attached OM Errata Sheet and save it with the Operations Manual you currently have available for future references regarding the "safety distance" required for pacemakers. 3. Please verify that the modules listed in Table 1 have the pacemaker safety label attached. If the safety pacemaker label is missing, please remove the last page of this letter and tape the copy of the label on the cover of these modules until the official label is available. If your site is included in the list of Systems where at least one pacemaker label is missing, lnpeco Service or their representatives will contact you to arrange a visit to address the described problem and apply the missing labels. In addition: " complete and return the Field Safety Notice Receipt Confirmation form attached to this letter within 15 days directly to lnpeco or to lnpeco representative. If you have any questions or concerns regarding this recall, please contact Giorgia Amabile, Quality and Regulatory Manager, at Regulatory.Affairs@inpeco.com, or (+41) 91-9118-258.
Initiated
2018-08-24
Root cause
Under Investigation by firm
Terminated
2020-03-19
Distribution
Foreign distribution to Netherlands, Italy, Belgium, Austria, France, Spain, and Germany.
Recall number
Z-0361-2019
Recall status
Terminated
CDRH record ID
167487
FDA record created
2018-10-31
Recall event number
81124

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗