Skip to content

Public FDA record

Z-0362-06: Roche brand Omni S4 Blood Gas Analyzer; Catalog # 03337138001.

Recall record Z-0362-06 for Roche Diagnostics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0362-06
Company in FDA record
Roche Diagnostics Corp.
FDA classification date
Jan 7, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Potential for hemoglobin derivatives which are outside of the quality control (QC) limits to not engage the auto lock out feature of the system for 02Hb, COHb, HHb, MetHb or SO2.

Source facts

Action
Consignees were notified via letter dated 12/9/05.
Initiated
2005-12-09
Root cause
Other
Terminated
2006-04-10
Distribution
California, Florida, Georgia, Indiana, Iowa, New Hampshire, Michigan, New York, North Carolina, North Dakota, Pennsylvania, Texas and Washington.
Recall number
Z-0362-06
Recall status
Terminated
CDRH record ID
43251
FDA classified
2006-01-07
Recall event number
34180

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗