Open original FDA record ↗Recall record
- FDA identifier
- Z-0362-2011
- Company in FDA record
- SCC Soft Computer
- FDA classification date
- Nov 15, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
SCC Soft Computer is issuing a Correction for SoftBank II Versions: 23.1.2.x, 23.2.0.x, and 25.0.0.x software applications. Functionality issue - For clients generating patient history reports using Patient>Reports>Batch Reports>History Report option in SoftBank with SoftReports activated for patient reports, when the report is generated for more than one patient ALL of the results for the qualif
Source facts
- Action
- SCC Soft Computer notified Customers on June 22, 2010, via the proprietary communication software entitled "Risk-to-Health Notification Task Verbiage" dated June 11, 2010. The notification described the product, problem and action to be taken by customer. The customers were instructed to acknowledge receipt of the notification; to grant permission to load the required hot fix, and to make the manual correction. For the manual correction to the datapath for patient report layouts, your SoftBank Support Specialist will Coordinate those changes and your validation of affected environments. If you have any questions, contact (727) 789-0100.
- Initiated
- 2010-06-22
- Root cause
- Software design
- Terminated
- 2012-09-05
- Distribution
- Worldwide distribution: U.S. and country of Canada.
- Recall number
- Z-0362-2011
- Recall status
- Terminated
- CDRH record ID
- 94757
- FDA classified
- 2010-11-15
- Recall event number
- 56872