Z-0362-2011: SoftBank Versions 23.1.2.x, 23.2.0.x, and 25.0.0.x with SoftReports. SCC Soft Computer 5400 Tech Data Dr

Recall record Z-0362-2011 for SCC Soft Computer. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0362-2011
Company in FDA record
SCC Soft Computer
FDA classification date
Nov 15, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

SCC Soft Computer is issuing a Correction for SoftBank II Versions: 23.1.2.x, 23.2.0.x, and 25.0.0.x software applications. Functionality issue - For clients generating patient history reports using Patient>Reports>Batch Reports>History Report option in SoftBank with SoftReports activated for patient reports, when the report is generated for more than one patient ALL of the results for the qualif

Source facts

Action
SCC Soft Computer notified Customers on June 22, 2010, via the proprietary communication software entitled "Risk-to-Health Notification Task Verbiage" dated June 11, 2010. The notification described the product, problem and action to be taken by customer. The customers were instructed to acknowledge receipt of the notification; to grant permission to load the required hot fix, and to make the manual correction. For the manual correction to the datapath for patient report layouts, your SoftBank Support Specialist will Coordinate those changes and your validation of affected environments. If you have any questions, contact (727) 789-0100.
Initiated
2010-06-22
Root cause
Software design
Terminated
2012-09-05
Distribution
Worldwide distribution: U.S. and country of Canada.
Recall number
Z-0362-2011
Recall status
Terminated
CDRH record ID
94757
FDA classified
2010-11-15
Recall event number
56872

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗