Skip to content

Public FDA record

Z-0362-2014: Fixed Offset Adapter, A-642100. The Offset Adapter is a fixed modular component which sets the tube 14.5

Recall record Z-0362-2014 for Ossur North America Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0362-2014
Company in FDA record
Ossur North America Inc
FDA classification date
Nov 19, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ossur inititated a voluntary recall of the A-642100 Fixed Offset Adapter because it can develop fatigue cracks in normal use over an extended period of time and ultimately separate completely.

Source facts

Action
Ossur sent an "Safety Notice Voluntary Product Recall" letter dated October 29, 2013, via regular mail. Follow-up telephones are also made. The letter identified the product the problem and the actions to be taken by the customer. ssur is committed to providing superior customer service and we will do our utmost to minimize any inconvenience this may cause your practice. To ensure patient safety we kindly ask you to take the following action: " Please identify the users of each of the A-642100 adapters, see attached list. " Also, please check your inventory for any A-642100s you may have on hand but have not fitted on users. A follow-up phone call will be made a few days later to coordinate replacements. A new replacement component will be provided to the customer. For further questions please call (949) 360-3634.
Initiated
2013-10-24
Root cause
Device Design
Terminated
2014-06-17
Distribution
US Distribution including the states of AL, TX, NY, CA, VA, GA, WI, NC, ID and NJ.
Recall number
Z-0362-2014
Recall status
Terminated
CDRH record ID
122724
FDA classified
2013-11-19
Recall event number
66609

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗