Z-0362-2017: Symbiq Two Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids. List numb

Recall record Z-0362-2017 for Hospira Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0362-2017
Company in FDA record
Hospira Inc.
FDA record creation date
Nov 4, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

May go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The infusion and the processor stop, sound a 3 minute audible alarm from the backup buzzer and deactivate the couch screen and the automated loading/unloading of cassettes. Delay or interruption of therapy.

Source facts

Action
Hospira sent an Urgent Device Recall Letters, dated February 5, 2013, to all affected customers. Customers were directed to notify their consignees if they further distributed the product and confirm they have done so by return of the reply form. Healthcare professionals were to contact Stericycle. Hospira discontinued support of these devices as of June 30, 2015. Customers with questions were instructed to call 1-800-201-9064. For questions regarding this recall call 224-212-4892.
Initiated
2011-03-16
Root cause
Device Design
Terminated
2016-11-28
Distribution
Worldwide Distribution - US (nationwide) and Canada
Recall number
Z-0362-2017
Recall status
Terminated
CDRH record ID
115203
FDA record created
2016-11-04
Recall event number
64005

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗