Skip to content

Public FDA record

Z-0362-2020: TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse I IP 20, UDI: 04260498580

Recall record Z-0362-2020 for Tec Com Gmbh. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0362-2020
Company in FDA record
Tec Com Gmbh
FDA record creation date
Nov 12, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The action is being initiated due to potential installation of faulty peltier-elements within the thermo-regulation system. Failure of the peltier-element would report a system failure and no longer be able to guarantee temperature control at full power.

Source facts

Initiated
2018-08-14
Root cause
Nonconforming Material/Component
Terminated
2019-12-20
Distribution
US Nationwide distribution in the state of UT.
Recall number
Z-0362-2020
Recall status
Terminated
CDRH record ID
176201
FDA record created
2019-11-12
Recall event number
83707

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗