Skip to content

Public FDA record

Z-0362-2023: IgA Immunoturbidimetric Assay- In vitro determination of IgA and IgM in serum and plasma Catalog Number:

Recall record Z-0362-2023 for Randox Laboratories Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0362-2023
Company in FDA record
Randox Laboratories Ltd.
FDA classification date
Dec 3, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Update to the carryover avoidance technical bulletin to detail that IgA and IgM assays should not be run immediately after Fructosamine Catalogue Number FR3133 or FR4030 on the RX instruments testing order. Can lead to an elevation in Quality Control and Patient results of up to +13% for lgA and +51% for lgM respectively, which may lead to a delay in running patient samples or erroneous elevated test results

Source facts

Action
Randox Laboratories extended recall notification to the US Distribution center on 11/01/22 via email. The US distributor contacted customers on 11/1/22 via email . Letter states reason for recall, health risk and action to take: Review your instrument testing order in line with the Carryover Avoidance Technical Bulletin (RXTB-0136} " Update the RX user manual with the updated Carryover Avoidance Technical Bulletin (RXTB-0136} and ensure all operators are aware of the recommendations. Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation. If you have any questions or concerns, please contact Randox Technical Services
Initiated
2022-11-01
Root cause
Under Investigation by firm
Terminated
2024-06-18
Distribution
US Nationwide Distribution: CA, NC, NJ and WV,
Recall number
Z-0362-2023
Recall status
Terminated
CDRH record ID
196724
FDA classified
2022-12-03
Recall event number
91097

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗