Open original FDA record ↗Recall record
- FDA identifier
- Z-0362-2025
- Company in FDA record
- Topcon Medical Systems, Inc.
- FDA classification date
- Nov 12, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The overlay of visual field test locations on a probability map, as presented in IMAGEnet 6 in the so-called "Hood Report", is not appropriate as it can imply a structure/function correlation for which clinical evidence is not provided
Source facts
- Action
- Topcon issued Product and Service Bulletin titled: Important notice regarding changes to IMAGEnet 6 for Maestro2 users on 9/17/24 via email. Action Required: In order to comply with recent FDA requirements, changes have been made to the Hood Report for Glaucoma on the Maestro2. You must visit our website to download the software patch for your device. This simple process is designed for customers to complete independently and should take less than 20 minutes to update. Download Software Patch For questions or assistance, please contact customer service at 1-866-922-6278 Mon Fri, 8AM-8PM EST
- Initiated
- 2024-09-17
- Root cause
- Software design
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0362-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210362
- FDA classified
- 2024-11-12
- Recall event number
- 95485