Z-0362-2025: The IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, s

Recall record Z-0362-2025 for Topcon Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0362-2025
Company in FDA record
Topcon Medical Systems, Inc.
FDA classification date
Nov 12, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The overlay of visual field test locations on a probability map, as presented in IMAGEnet 6 in the so-called "Hood Report", is not appropriate as it can imply a structure/function correlation for which clinical evidence is not provided

Source facts

Action
Topcon issued Product and Service Bulletin titled: Important notice regarding changes to IMAGEnet 6 for Maestro2 users on 9/17/24 via email. Action Required: In order to comply with recent FDA requirements, changes have been made to the Hood Report for Glaucoma on the Maestro2. You must visit our website to download the software patch for your device. This simple process is designed for customers to complete independently and should take less than 20 minutes to update. Download Software Patch For questions or assistance, please contact customer service at 1-866-922-6278 Mon Fri, 8AM-8PM EST
Initiated
2024-09-17
Root cause
Software design
Distribution
US Nationwide distribution.
Recall number
Z-0362-2025
Recall status
Open, Classified
CDRH record ID
210362
FDA classified
2024-11-12
Recall event number
95485

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗