Open original FDA record ↗Recall record
- FDA identifier
- Z-0363-2008
- Company in FDA record
- Blackstone Medical, Inc.
- FDA classification date
- Dec 11, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Manufacturing Defect: The saddle of the screw's body relative to the keyway exceeds the tolerance outlined in the device specifications. The extent to which the tolerance is exceeded varies from part to part, in the extreme the tolerance is exceeded by (+) 50%. As a consequence, there is a potential for set-screw cross-threading during implantation. Cross-threading could in turn lead to a comprom
Source facts
- Action
- Blackstone Medical, Inc., field representatives were notified via phone, e-mail or fax on 4/30/07. Attention: Notification of Product Removal Letters were sent on May 15, 2007, to distributors and implanting surgeons requesting return of non-implanted stock.
- Initiated
- 2007-05-15
- Root cause
- Nonconforming Material/Component
- Terminated
- 2008-06-06
- Distribution
- Worldwide: USA and countries of China, Cyprus, Turkey and Switzerland.
- Recall number
- Z-0363-2008
- Recall status
- Terminated
- CDRH record ID
- 65731
- FDA classified
- 2007-12-11
- Recall event number
- 45523