Open original FDA record ↗Recall record
- FDA identifier
- Z-0363-2012
- Company in FDA record
- SCC Soft Computer
- FDA classification date
- Dec 2, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
On 07/20/2011 SCC Soft Computer, Clearwater, FL, initiated a correction on the following SoftPath GUI versions: 4.2.3.3-4.2.3.6 and 4.3.0.5-4.3.0.12. A client reported that the HPV results on a patient report were incorrect. The HPV results were from the previous order on the patient and not the current order. All affected clients were notified about the issue on 07/20/2011.
Source facts
- Action
- SCC Soft Computer sent a "CORRECTION COMMUNICATION" dated July 20, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. SCC personnel will review the customized Lab Results AP macro and/or the Micro Results AP macro on customer systems and will either discuss with them the irregularities requiring change or reconfigure macros as necessary.
- Initiated
- 2011-07-20
- Root cause
- Software design
- Terminated
- 2013-05-21
- Distribution
- Worldwide Distribution-USA (nationwide) and the country of Canada.
- Recall number
- Z-0363-2012
- Recall status
- Terminated
- CDRH record ID
- 105445
- FDA classified
- 2011-12-02
- Recall event number
- 60458