Skip to content

Public FDA record

Z-0363-2024: Ventlab, LLC Horizon manual resuscitators with integrated manometer, multiple accessory configurations to

Recall record Z-0363-2024 for SunMed Holdings, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0363-2024
Company in FDA record
SunMed Holdings, LLC
FDA classification date
Nov 21, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

Source facts

Action
Customer notification was emailed October 11, 2023 advising customers that monitoring of tidal carbon dioxide in critically ill patients is recommended because use of accessories (e.g. PEEP valve) on the exhalation port of manual resuscitators may increase backward leak in the integrated manometer due to increased expiratory resistance. This affects all lots with duckbill-style patient valve with integrated manometer manufactured since 09/21/2019. Customers are to disseminate this information to all users and complete and return the response for. The notice also notes manual resuscitator IFUs will be updated to include this information as a warning statement.
Initiated
2023-10-11
Root cause
Device Design
Distribution
US Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.
Recall number
Z-0363-2024
Recall status
Open, Classified
CDRH record ID
203764
FDA classified
2023-11-21
Recall event number
93237

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗