Z-0363-2025: Irrigation Tubing for KARL STORZ Endomat Select UP210 REF 031523-10 Tubing Set, Irrigation, PC

Recall record Z-0363-2025 for Karl Storz Endoscopy. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0363-2025
Company in FDA record
Karl Storz Endoscopy
FDA classification date
Nov 12, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to improper labeling of products. An intended use on the label has not been reviewed and approved by FDA for distribution in the U.S.

Source facts

Action
On 10/22/2024, an "URGENT: MEDICAL DEVICE CORRECTION" Letters will be sent via email by KARL STORZ informing customers that the firm has become aware that one lot number of misbranded product was distributed to 19 customers. Customers are instructed to: 1. Pass on this Urgent Medical Device Correction to all users of the products listed above (Table 1) and all other persons who need to be aware within your organization. 2. Communicate to all applicable users that the product is not to be utilized during spinal procedures. 3. If you have or may have distributed the products listed, please identify and promptly notify those recipients of this notice. 4. Return the completed Customer Reply Form by E-Mail to the indicated contact below within 15 calendar days from the date of receipt. 5. Please report any incidents related to this issue to the manufacturer, dealer, or local representative and, if applicable, to the national competent authority, as this is important feedback. for questions send an email to KSEAProductLabeling@karlstorz.com .
Initiated
2024-10-21
Root cause
Labeling design
Distribution
U.S. Nationwide distribution in the states of CO, CT, DC, MD, MI, MN, NC, NY, OH, OK, TN, TX, WA.
Recall number
Z-0363-2025
Recall status
Open, Classified
CDRH record ID
210760
FDA classified
2024-11-12
Recall event number
95607

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗