When using the Creatine Clearance dose calculations a software bug will allow the algorithm to assume a patients age to be in years when it is entered in months. This miscalculation may result in patients less than two years old to be seriously overdosed with radiation.
Source facts
Action
The firm has sent out notification letters to the 18 sites with the activated MedOnc option. All 18 site had the letters delivered by Siemens Medical Solutions field service personnel by 1/6/2005.
Initiated
2005-01-06
Root cause
Software design
Terminated
2006-05-15
Distribution
A total of 18 sites were identified by the firm as having the MedOnc module activated; 6 within the United States U.S.) and 12 outside of the U.S.
Recall number
Z-0364-06
Recall status
Terminated
CDRH record ID
43217
FDA classified
2006-01-07
Recall event number
34166