Open original FDA record ↗Recall record
- FDA identifier
- Z-0364-2008
- Company in FDA record
- Accuray Inc
- FDA classification date
- Dec 12, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Software problem - Sagittal, oblique or coronal sliced MR Images imported from a non-Accuray manufactured device into the product may result in inverted image orientation.
Source facts
- Action
- The firm issued an Urgent Advisory Notification on March 20, 2007, notifying users to only import axial sliced MR image series for treatment planning. Users should not import sagittal, oblique or coronal images series until further notice. The firm intends to issue a software patch to correct the problem and will contact user regarding resolution. Users can contact Accuray Customer Support at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or customersupport@accuray.com.
- Initiated
- 2007-03-21
- Root cause
- Software design
- Terminated
- 2007-12-26
- Distribution
- Worldwide Distribution - USA including states of MD, CA, FL, WA, LA, IL, IN, MI, MN, NC, NJ, OK, WI, and OH, and countries of Korea, Greece, Germany, France, and Turkey.
- Recall number
- Z-0364-2008
- Recall status
- Terminated
- CDRH record ID
- 53465
- FDA classified
- 2007-12-12
- Recall event number
- 38346