Open original FDA record ↗Recall record
- FDA identifier
- Z-0364-2011
- Company in FDA record
- SCC Soft Computer
- FDA classification date
- Nov 15, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Functionality issue: For clients using the instrument to generate results on the interface menu, two different result records could be displayed under the same order and sequence numbers. This could cause incorrect results to be populated to a patient record.
Source facts
- Action
- Clients were notified via the proprietary communication software on 06/01/2010. The communication discussed the risk-to-health notification task, potential hazard, root cause, corrective action, and requested action. Clients were instructed to acknowledge receipt of the notification. If the client wishes to receive the correction, they should document a request to load the software.
- Initiated
- 2010-06-01
- Root cause
- Software design
- Terminated
- 2012-07-16
- Distribution
- Worldwide Distribution -- US, including states of CT, FL, IL, KY, GA, LA, IN, MS, NJ, NY, PA, OH, and TN and country of Canada.
- Recall number
- Z-0364-2011
- Recall status
- Terminated
- CDRH record ID
- 94685
- FDA classified
- 2010-11-15
- Recall event number
- 56865