Skip to content

Public FDA record

Z-0364-2014: Juggerknotless Drill Bit, REF 110003173, Sterile, Single Use Product Usage: The Juggerknotless Soft Anc

Recall record Z-0364-2014 for Biomet, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0364-2014
Company in FDA record
Biomet, Inc.
FDA classification date
Nov 19, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Investigation identified that the design of the drill bit may make the suture anchor difficult to insert into the bone. The implant may skip across the hole or miss the hole during insertion.

Source facts

Action
Biomet sent an Urgent Medical Device Recall Notice letter dated October 10, 2013 to all consignees. The letter identified the affected product, reason for recall, and actions to be taken. The letter stated customers will be contacted on an individual basis in an attempt to reconcile the affected product. Customers were directed to locate and remove implicated devices and fax a copy of the response form to 574-372-1683 prior to returning product. For questions call 574-372-1570.
Initiated
2013-10-10
Root cause
Device Design
Terminated
2014-07-16
Distribution
USA Nationwide Distribution in the states of VA, CA, Ky, and IN
Recall number
Z-0364-2014
Recall status
Terminated
CDRH record ID
122873
FDA classified
2013-11-19
Recall event number
66639

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗