Z-0364-2023: FULL VISION INC. TRACKMASTER, TREADMILL TMX428 110V, Model #317-07926

Recall record Z-0364-2023 for Full Vision Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0364-2023
Company in FDA record
Full Vision Inc
FDA classification date
Dec 3, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.

Source facts

Action
The firm began sending URGENT MEDICAL DEVICE CORRECTION notices its consignees by email beginning 09/13/2022. The distributors were provided with a letter to distribute to their customers (end users) The notice explained the issue, risk to health, and requested the following actions be taken: "Actions to be taken by the Customer/User: A simple test can be performed to determine if the unit has the incorrect component installed on the PCB Assembly. The user should turn the device on. Then cycle the power with the power switch by turning it off, waiting 5 seconds, then turn it back on. Start the belt at 2 mph and pull the Magnetic Tether. If the assembly has the incorrect component the device should fail to stop. If the assembly has the correct component the device should stop. The assemblies that fail to stop should be replaced with the c-clip style safety tethers, provided by Full Vision to the distributors and end users. To continue using the treadmill with the Magnetic Tether assemblies until they can be replaced, you can prevent this condition from occurring by cycling the power for 30 seconds before turning the device back on. Product Correction: Full Vision has scrapped all the non-conforming components (Magnetic Safety Tether and PCB Board Assembly) from the inventory. All nonconforming components will be replaced."
Initiated
2022-08-31
Root cause
Other
Distribution
Wi, GA, NY, Mexico
Recall number
Z-0364-2023
Recall status
Open, Classified
CDRH record ID
196451
FDA classified
2022-12-03
Recall event number
90995

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗