Skip to content

Public FDA record

Z-0364-2025: Brand Name: UroPass Ureteral Access Sheaths Product Name: Ureteral Access Sheaths Model/Catalog Number: 6

Recall record Z-0364-2025 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0364-2025
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Nov 12, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Exposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures.

Source facts

Action
****UPDATE Nov 8, 2024*** Olympus mailed a revision letter-URGENT: MEDICAL DEVICE CORRECTION UPDATED INFORMATION to consignees on November 8, 2024. Letter states; To reduce the risk of UV exposure to your device(s), Olympus instructs users to implement the following actions: Minimize the exposure of UroPass Ureteral Access Sheaths to ultraviolet light. Exposure to UV light during storage periods may cause embrittlement of this device, increasing the potential for breakage and patient injury, including retained device fragments, tissue damage, or perforation of the ureter. Updated Information: " As instructed in IFU Section 4.0, inspect the device prior to use. If any damage to the device is noted, DO NOT USE. " Inspect the device upon removal or completion of the procedure. If damage is noted, the patient should be evaluated for a potentially retained foreign body. " Sharp edges of a broken sheath have been reported to have caused ureteral tissue damage or perforation. If device damage is noted, patients should be monitored intraoperatively and in the immediate post-operative period for these types of injuries. Olympus is not requiring the return of your UroPass device(s) as a result of this action. However, if you want to return the UroPass device(s) in your inventory, please contact Customer Service at 1-800-848-9024, option 2 to obtain a Return Material Authorization. Olympus will issue a credit to your facility upon return of your affected product. 4. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification, including the new information regarding UV exposure. Olympus is in the process of updating the Instructions for Use with this information. 5. Olympus requests that you acknowledge receipt of this letter. Access the Olympus recall portal to indicate that you have received this notification. a. Go to https://olympusamerica.com/recall b. Enter the recall number "0458 c. Complete the form as instructed. 6. If
Initiated
2024-09-11
Root cause
Storage
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Canada, China, Germany, Hong Kong, Japan, Republic of South Korea, Singapore.
Recall number
Z-0364-2025
Recall status
Open, Classified
CDRH record ID
210637
FDA classified
2024-11-12
Recall event number
95570

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗