Z-0366-2012: Zimmer ZMR Femoral Spout & Cone Bodies and Trabecular Metal Primary Hip Prosthesis Femoral Stem SIZE 16

Recall record Z-0366-2012 for Zimmer Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0366-2012
Company in FDA record
Zimmer Inc.
FDA record creation date
Jun 26, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Lack of assurance of proper metal fatigue strength due to metal grain structure anomaly. In 2006 a recall was initiated as a result of the grain anomaly; however, these 4 lots had been split off during normal production and had not been included in the 2006 recall.

Source facts

Action
On 11/16/11, Zimmer sent letters titled "URGENT MEDICAL DEVICE RECALL" to distributors via electronic mail and to risk managers and surgeons via FedEx indicating that the product must be immediately removed from use and quarantined. " Stop using the affected lots and place affected product in quarantine. " Review available product on the shelf " Identify those part/lot combinations listed in the recall communication " Immediately remove any implants found that match part/lot combinations listed in the recall communication and quarantine " Return affected lots and signature of acknowledgement to Zimmer
Initiated
2011-11-16
Root cause
Nonconforming Material/Component
Terminated
2013-08-28
Distribution
Worldwide distribution: USA (NY), AUSTRALIA, SWITZERLAND, and DENMARK
Recall number
Z-0366-2012
Recall status
Terminated
CDRH record ID
105374
FDA record created
2015-06-26
Recall event number
60419

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗