Open original FDA record ↗Recall record
- FDA identifier
- Z-0366-2020
- Company in FDA record
- TriMed Inc.
- FDA record creation date
- Nov 13, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm has become aware that the angled-hole of certain semi-tubular design bone plates may have been modified from its specification, and the modification may not have been fully qualified.
Source facts
- Action
- On 08/15/2019 and 9/13/19, the firm issued a recall notice to affected consignees via email. The recall notification informed consignees of the following: 1. Review your inventory to see if you have any of the impacted product. If you have impacted product within your inventory, remove the product use. Do not implant the impacted product. Contact TriMed at quality@trimedortho.com or call 800-633-7221. TriMed will arrange to have the impacted product returned to the company. 2. Review your records to see if the impacted product was implanted. If you have records that impacted product was implanted, medical judgment should be applied to determine whether additional patient monitoring activities should be performed to evaluate implant placement, implant structural integrity, and fracture union. 3. Return the Recall Response card attached to this letter to TriMed as soon as possible. 4. Please contact TriMed via telephone or email as per below if you have any questions:, Telephone*: 800-633-7221 Email: quality@trimedortho.com, Hours of operation: 8:00AM to 5:00PM PST (Monday through Friday, excluding holidays)
- Initiated
- 2018-08-15
- Root cause
- Employee error
- Distribution
- US Nationwide distribution in the states of MD, UT.
- Recall number
- Z-0366-2020
- Recall status
- Terminated
- CDRH record ID
- 176904
- FDA record created
- 2019-11-13
- Recall event number
- 83984