Z-0366-2026: Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 1038144

Recall record Z-0366-2026 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0366-2026
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Oct 27, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on the IMMULITE 2000/IMMULITE 2000 XPi systems. The bias is observed in both serum and plasma samples.

Source facts

Action
On September 22, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer Actions: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Discontinue use of and discard the kit lot listed in the table above (Products Section). Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. For continued Intact PTH testing in your laboratory during this time, please contact your Siemens representative to discuss alternative Siemens Healthineers solutions. As an alternative, customers who have an ADVIA Centaur system or an Atellica IM/CI may utilize the corresponding iPTH assay on those systems. These platforms are not impacted. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. A follow-up communication will be provided when assay performance has been restored. ***Update 2/18/2026*** On February 10, 2026 Follow-up Urgent Medical Device Correction letters were sent to additional customers to notify of the addition of new lots for the recall. Customer actions remained the same as the original letter.
Initiated
2025-09-22
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, DC, FL, ID, IN, LA, MD, NC, NJ, NY, PA, SC and the countries of Argentina, Australia, Bolivia, Bosnia Herzeg, Brazil, Bulgaria, Canada, Chile, Colombia, Cyprus, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, Hungary, India, Iran, Israel, Italy, Kosovo, Kuwait, Latvia, Lebanon, Macedonia, Mexico, Nepal, Netherlands, P.R. China, Pakistan, Palestinian Ter, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Serbia, Slovakia, Slovenia, South Africa, Spain, Taiwan, Tunisia, Turkey, Turkmenistan, U.A.E., Ukraine, Uruguay, Uzbekistan.
Recall number
Z-0366-2026
Recall status
Open, Classified
CDRH record ID
216317
FDA classified
2025-10-27
Recall event number
97672

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗