Z-0367-05: System 1000 family of Hemodialysis Instruments, including the System 1000, Arena, 1550, BM11, BM25 and Me

Recall record Z-0367-05 for Baxter Healthcare Renal Div. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0367-05
Company in FDA record
Baxter Healthcare Renal Div
FDA classification date
Dec 24, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Air bubbles have been observed in the tubing sets past the air detector of the instrument with no alarms. Air bubbles in the circuit can cause an air embolism.

Source facts

Action
An Urgent Device Correction letter dated 12/14/04 was sent along with a reply form and addenda to operator''s manuals to all Baxter customers who have purchased hemodialysis hardware. The letter informed the accounts of reports of air bubbles observed in the tubing sets past the air detector without an alarm alerting the operator of potential air in the tubing. A review of the reports revealed misuse, including failure to follow generally observed hemodialysis practices. Baxter updated their Hemodialysis Operator''s Manuals and Training Guides to clarify instructions and add caution and warning statements to address the possibility of introducing air in the extracorporeal circuit and to clarify methods for clearing air from the circuit. The accounts were requested to review the attached addendum that applies to the instruments used in their facility against the pertinent sections of their Operator''s Manual. If their Operator''s Manual does not match the addendum, they were instructed to order a new manual using the enclosed response form. Any questions were directed to Baxter''s Global Technical Services at 1-800-553-6898.
Initiated
2004-12-14
Root cause
Other
Terminated
2008-02-26
Distribution
This action covers Baxter''s complete line of hemodialysis hardware, which was distributed nationwide, including Puerto Rico, and internationally to Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Denmark, Ecuador, Finland, France, Republic of Georgia, Greece, Guatemala, Honduras, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Korea, Lebanon, Morocco, Malaysia, Netherlands, Oman, Panama, Peru, Philippines, Poland, Palestine, Romania, Russia, Saudia Arabia, Singapore, Slovakia, Spain, Sweden, Taiwan, Thailand, Tunisia, Turkey and the United Kingdom.
Recall number
Z-0367-05
Recall status
Terminated
CDRH record ID
36019
FDA classified
2004-12-24
Recall event number
30507

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗