Z-0367-2020: RANDOX NEFA (Non-Esterified Fatty Acids) FA115 Lot 485343, GTIN: 5055273203066.

Recall record Z-0367-2020 for Randox Laboratories, Limited. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0367-2020
Company in FDA record
Randox Laboratories, Limited
FDA record creation date
Nov 13, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Non-Esterified Fatty Acids (NEFA) FA115 batch 485343 is not meeting the performance claims and should not be used in any further testing.

Source facts

Action
On September 23, 2019 the firm distributed Urgent Medical Device Recall letters by e-mail instructing them to: - Discontinue use of and discard any of the product detailed above immediately or return remaining product to Randox Laboratories. - Review results generated with the affected batches in line with the clinical profile of the patient. - Discuss the contents of this notice with your Medical Director. - Review your reagent inventory of these products and assess your laboratories needs for reimbursement for discarded inventory. - Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organization. On October 8, 2019 the firm distributed an amended Urgent Medical Device Recall. However, these letters were not identified as an Amended or Corrected notice. The notice is also still dated 09/23/2019. This updated notice identified the expiration date for the affected product as 28Mar2021. On October 16, 2019, the firm sent a second amended notice to customers via email.
Initiated
2019-09-23
Root cause
Nonconforming Material/Component
Terminated
2020-05-07
Distribution
WV (USA), International: GB, BE, CZ, FR, GR, IE, IT, NL, PL, SK, ES, CH, TR, AU, BR, MX, RU, SG, TW
Recall number
Z-0367-2020
Recall status
Terminated
CDRH record ID
176991
FDA record created
2019-11-13
Recall event number
84007

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗