Open original FDA record ↗Recall record
- FDA identifier
- Z-0367-2022
- Company in FDA record
- Mentor Texas, LP.
- FDA record creation date
- Dec 10, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a manufacturing issue, Implant may have a potentially weakened area on the base patch portion of the implant which may cause leakage/deflation of the affected device.
Source facts
- Action
- On 10/11/2021, the Firm sent an "URGENT MEDICAL DEVICE RECALL (REMOVAL)" Notfication via UPS informing customers that breast implants which have expiration dates from 01/01/2025 to 09/30/2021, due to a manufacturing issue may have a potentially weakened area on the base patch portion of the shell. As a result of this issue, these implants may prematurely deflate for which replacement surgery may be needed. Customers are instructed to: 1. Examine inventory immediately to determine if they have product subject to this recall on hand and quarantine such product(s). 2. Remove the product subject to this recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in their facility who needs to be informed. 3. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Consider including a copy of this recall letter when communicating. 4. Complete the Business Reply Form (BRF) (Attachment 2) confirming receipt of this notice and fax/email to Sedgwick at 888-912-7086 or mentor7209@sedgwick.com within 3 business days. Return the BRF even if they do not have product subject to this recall. 5. Follow instructions in the letter and immediately return any inventory of MENTOR Smooth Round Saline DV Breast Implants with expiration dates ranging from 01/01/2025 to 09/30/2025 subject to this recall to Sedgwick. 6. Keep this notice visibly posted for awareness until all product subject to this recall has been returned to Sedgwick. While processing the returns, maintain a copy of this notice with the product subject to this recall and keep a copy for your records. 7. Return unused impacted Breast Implants with expiration dates ranging from 01/01/2025 to 09/30/2025 subject to this recall that are in their inventory immediately. To receive credit reimbursement, customers must return product subject to this recall. Any non-affected product and any product returned after the dat
- Initiated
- 2021-09-30
- Root cause
- Process control
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Canada, Spain, UK, Germany, France, Korea, Australia, Taiwan, Belgium, Ireland, Turkey, and Singapore.
- Recall number
- Z-0367-2022
- Recall status
- Open, Classified
- CDRH record ID
- 189789
- FDA record created
- 2021-12-10
- Recall event number
- 88868