Skip to content

Public FDA record

Z-0367-2026: Various models of BD Pyxis Medbank: Reference numbers: 1137-00 1145-00 1146-00 1147-00 1148-00 138902-0

Recall record Z-0367-2026 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0367-2026
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Oct 27, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.

Source facts

Action
On September 30, 2025, BD issue a Urgent Medical Device Recall Expansion notification to affected consignees via FedEx. BD is asking consignees to take the following actions: 1. Affected products can continue to be used per existing Manuals and Guides. 2. In the event of an accidental fluid spill on the device or in a drawer, users should following instructions outlined within the customer notification. 3. Review the customer letter attachment which provides warnings, cautions and recommendations regarding fluid ingress. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material.
Initiated
2025-09-30
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, United Arab Emirates, Australia.
Recall number
Z-0367-2026
Recall status
Open, Classified
CDRH record ID
215942
FDA classified
2025-10-27
Recall event number
97566

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗