Skip to content

Public FDA record

Z-0368-04: Folate Low and High Controls, list 9C13-10; the pack contains 8 - 2 mL bottles( 4 for each level), Low Co

Recall record Z-0368-04 for Abbott Laboratories HPD/ADD/GPRD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0368-04
Company in FDA record
Abbott Laboratories HPD/ADD/GPRD
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The is an increase in the number of control values out of range low, and diluted patient specimens may yield lower than expected results.

Source facts

Action
Abbott notified their international affiliates via e-mail on 12/16/03 and 1/5/04. Device recall letters dated 12/16/03 and 12/31/03 will be sent to the end users by the affiliates, informing them of the unapproved serum albumin used to manufacture the folate products, and listing the impact to performance for each affected components. There are separate customer recall letters for each assay platform -- ARCHITECT, AxSYM and IMx.
Initiated
2003-12-16
Root cause
Other
Terminated
2004-07-20
Distribution
There was no U.S. distribution. The products were shipped internationally through Abbott affiliates in Canada, Germany, Singapore, Hong Kong, Australia, Brazil, New Zealand, Columbia, Chile, Uruguay, United Kingdom, Venezuela, Dominican Republic, Argentina and Curacao.
Recall number
Z-0368-04
Recall status
Terminated
CDRH record ID
31085
FDA classified
2004-07-20
Recall event number
27978

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗