Possible overexposure: Failure to apply the entrance exposure rate (EER) tube current limit calibration to the automatic exposure rate (AER) control system when the anatomical profile mode is changed from the default selection to another selection. Manual technique modes are not affected.
Source facts
Action
GE OEC sent a letter to customers December 15, 2007, and scheduled field representative visits for corrective software solution installation.
Initiated
2007-12-14
Root cause
Software design
Terminated
2012-04-13
Distribution
Worldwide distribution: USA, Finland, Netherlands, Germany, Switzerland, France, United Kingdom, Australia, Italy, Ireland, Spain, Sweden, Belgium, New Zealand, China, Singapore, Poland, Norway, and India.
Recall number
Z-0368-2008
Recall status
Terminated
CDRH record ID
73760
FDA classified
2008-09-29
Recall event number
49536