Open original FDA record ↗Recall record
- FDA identifier
- Z-0368-2012
- Company in FDA record
- Zimmer Inc.
- FDA record creation date
- Jun 26, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Lack of assurance of proper metal fatigue strength due to metal grain structure anomaly. In 2006 a recall was initiated as a result of the grain anomaly; however, these 4 lots had been split off during normal production and had not been included in the 2006 recall.
Source facts
- Action
- On 11/16/11, Zimmer sent letters titled "URGENT MEDICAL DEVICE RECALL" to distributors via electronic mail and to risk managers and surgeons via FedEx indicating that the product must be immediately removed from use and quarantined. " Stop using the affected lots and place affected product in quarantine. " Review available product on the shelf " Identify those part/lot combinations listed in the recall communication " Immediately remove any implants found that match part/lot combinations listed in the recall communication and quarantine " Return affected lots and signature of acknowledgement to Zimmer
- Initiated
- 2011-11-16
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-08-28
- Distribution
- Worldwide distribution: USA (NY), AUSTRALIA, SWITZERLAND, and DENMARK
- Recall number
- Z-0368-2012
- Recall status
- Terminated
- CDRH record ID
- 105389
- FDA record created
- 2015-06-26
- Recall event number
- 60419