Skip to content

Public FDA record

Z-0369-2008: GE Precision 500D Classical Radiographic-Fluoroscopic X-ray System. Model #2289299-2, with console compu

Recall record Z-0369-2008 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0369-2008
Company in FDA record
GE Healthcare
FDA classification date
Sep 4, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

It was reported that when the GE Precision 500D foot pedal is repeatedly activated, a defect in the fluoro-timer will cause the dose measurement to report a higher value than was actually received by the patient. Patient will receive a lower dosage irradiation time which will result in indication of cumulative air kerma higher than that to which the patient was actually exposed.

Source facts

Action
A customer letter "GE Healthcare Product Safety Notification" dated 10/4/07 was distributed to customers informing them of the safety issue. A Field Modification Instructions (FMI 10806) was issued to the field for the Precision 500D product to install a sofware upgrade to correct the issue. This correction was released in October 2007.
Initiated
2007-10-04
Root cause
Radiation Control for Health and Safety Act
Terminated
2012-08-09
Distribution
AL, AR, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KY, LA, MA, MI, MN, MS, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, ITALY, Kuwait, Great Britan, Spain and Canada.
Recall number
Z-0369-2008
Recall status
Terminated
CDRH record ID
67914
FDA classified
2008-09-04
Recall event number
46680

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗