Open original FDA record ↗Recall record
- FDA identifier
- Z-0369-2008
- Company in FDA record
- GE Healthcare
- FDA classification date
- Sep 4, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
It was reported that when the GE Precision 500D foot pedal is repeatedly activated, a defect in the fluoro-timer will cause the dose measurement to report a higher value than was actually received by the patient. Patient will receive a lower dosage irradiation time which will result in indication of cumulative air kerma higher than that to which the patient was actually exposed.
Source facts
- Action
- A customer letter "GE Healthcare Product Safety Notification" dated 10/4/07 was distributed to customers informing them of the safety issue. A Field Modification Instructions (FMI 10806) was issued to the field for the Precision 500D product to install a sofware upgrade to correct the issue. This correction was released in October 2007.
- Initiated
- 2007-10-04
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2012-08-09
- Distribution
- AL, AR, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KY, LA, MA, MI, MN, MS, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, ITALY, Kuwait, Great Britan, Spain and Canada.
- Recall number
- Z-0369-2008
- Recall status
- Terminated
- CDRH record ID
- 67914
- FDA classified
- 2008-09-04
- Recall event number
- 46680