Open original FDA record ↗Recall record
- FDA identifier
- Z-0370-2010
- Company in FDA record
- Zimmer Inc.
- FDA classification date
- Nov 19, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The spacer cutter may jam during surgery, preventing use of the instrument and causing surgical delay.
Source facts
- Action
- Consignees were notified by letter dated 5/20/09 titled "Urgent Medical Device Correction" that instructed customers that they may continue to use the recalled product until adequate replacements become available, but that they should always ensure they have a backup spacer cutter (of a different model which does not have this issue) available in case this model should jam during use. Users were also advised to ensure the spacer cutter is properly lubricated and is sharp. For questions, please contact Zimmer Spine at 1-800-777-7505.
- Initiated
- 2009-05-28
- Root cause
- Other
- Terminated
- 2011-10-05
- Distribution
- Nationwide, Austria, Belgium, Canada, Italy, Netherlands, Singapore, Spain, Switzerland and United Kingdom.
- Recall number
- Z-0370-2010
- Recall status
- Terminated
- CDRH record ID
- 81682
- FDA classified
- 2009-11-19
- Recall event number
- 51933