Ergosafe Products LLC DBA Prism Medical Services USA
FDA classification date
Nov 20, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
The lift might fail with the labeled 500-lb. load rating.
Source facts
Action
The firm, Medcare Products, sent an "URGENT FIELD CORRECTIVE ACTIONS" letter dated October 21, 2013 via certified mail on October 30, 2013 to its customers. The letter described the product, problem and action to be taken. The customers were instructed to peel off the boom stickers and serial number label and replace with the new ones supplied; review the revised Manual and dispose of the original; and complete the action report form once the products have been relabeled. Please refer to your Operations Manual or visit our website at www.medcarelifts.com to view the manual online.
If we can answer any questions that you may have regarding the testing standards or the relabeling of the products, please do not hesitate to contact us at 1-800-695-4479.
Initiated
2013-10-30
Root cause
Device Design
Terminated
2017-05-17
Distribution
Nationwide distribution.
Recall number
Z-0371-2014
Recall status
Terminated
CDRH record ID
121498
FDA classified
2013-11-20
Recall event number
66056
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.