Z-0371-2025: The EVSRF catheter is a sterile, single-use disposable medical device for endovenous radiofrequency ablat

Recall record Z-0371-2025 for Bard Peripheral Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0371-2025
Company in FDA record
Bard Peripheral Vascular Inc
FDA classification date
Nov 13, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead to power being applied through the proximal 3/4th of the heating coil, bypassing the thermocouple, which will then cause the generator to overdrive the catheter in an attempt to reach desired temperature set point. The wire attachment error causes the device to reach temperatures exceeding the levels intended for therapeutic efficacy, while the console is falsely indicating that temperature has not yet been reached, thus increasing the risk of harm due to excessive temperature.

Source facts

Action
On 10/16/2024, the firm sent an "URGENT: Medical Device Product Removal" letter via FedEx or email to customers informing them that the Venclose RF Ablation Catheters may have been manufactured with internal wiring errors that could cause inaccurate temperature detection resulting in devices reaching temperatures higher than intended, while the user receives incorrect warnings that the catheter has not reached temperature. Customers are instructed to: 1. Please cease use of product listed in this letter, and immediately quarantine all affected product remaining in their possession. 2. The Venclose" RF Generator, which is used in conjunction with the Venclose" RF Ablation Catheters, contains software that requires an upgrade to be performed prior to using the impacted Venclose" RF Ablation Catheters. Venclose" RF Ablation Catheters that are not listed as impacted as per attachment 1, can continue to be used. 3. The affected devices should not be used until after the Venclose" RF Generator software has been updated by a BD representative. 4. A BD Representative will contact your facility in order to arrange the upgrade to your facility s Venclose" RF Generator software. The upgrade will be performed by a BD Representative upon completion of the Customer Response Form. 5. After the confirmed Venclose" RF Generator software has been updated, the products listed in this letter can be used as normal. 6. Share this notification with all users within your facility network of the product to ensure they are also aware of this field action. If the affected products have been distributed further, please ensure that these entities are contacted and informed of the recall. 7. Complete and return the attached Customer Response Form and return it to the BD contact noted on the form whether or not you have any of the impacted product so that BD may acknowledge your receipt of this notification. Indicate on the response form that the impacted product has been was quarantined by
Initiated
2024-09-04
Root cause
Employee error
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI and WV. The countries of Australia, Singapore, Austria, Belgium, Czechia, Denmark, France, Germany, Greece, Ireland, Italy, Poland, Slovenia Spain, Sweden, Switzerland, United Kingdom, United Arab Emirates, and South Africa.
Recall number
Z-0371-2025
Recall status
Open, Classified
CDRH record ID
210146
FDA classified
2024-11-13
Recall event number
95408

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗