Difficult to Close -- Certain lots of i-STAT cartridges are difficult to close, will not close or do not stay closed.
Source facts
Action
Abbott Point of Care sent and Urgent Notice to all i-STAT Customers (US and Canadian consignees) notifying them that certain lots of i-STAT cartridges have snap closures that are difficult to close, will not close or do not stay closed. The customers were asked to review the lot numbers on the enclosure and, if the facility has any cartridges from the lot numbers listed, they should remove from service immediately.
Initiated
2007-10-18
Root cause
Packaging change control
Terminated
2008-07-18
Distribution
Worldwide Distribution- including USA and country of Canada.
Recall number
Z-0372-2008
Recall status
Terminated
CDRH record ID
64692
FDA classified
2007-12-11
Recall event number
44972