- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Reports that the Rectal/ Esophageal Temperature Probes are not reading correctly. This situation causes the Blanketrol device, which the probes connect to, to shut down operation and display a 'check probe' error.
Source facts
- Action
- Cincinnati Sub-Zero Products, Inc., issued an "Urgent Medical Device Field Action" letter on November 1, 2011 to all affected customers. The letter includes; description of product and problem, and a list of the affected lot numbers. The letter instructs the customers to discontinue use of the device. All unused product is to be returned to the Biomedical/Clinical Engineering and/or Service Department for an replacement and all used probes are to be disposed of in accordance with hospital policy. The customers are also instructed to complete and return an attached Recall Response Form ASAP. For additional information, customers may contact the Beatrice Washington at 1-800-989-7373 or (513) 772-8810.
- Initiated
- 2011-09-26
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-08-13
- Distribution
- Nationwide distribution and Canada.
- Recall number
- Z-0372-2012
- Recall status
- Terminated
- CDRH record ID
- 105007
- FDA classified
- 2012-01-11