Skip to content

Public FDA record

Z-0372-2017: Replacement floor locking cylinders for the Schaerer model 7300 surgical tables. Replacement cylinders ar

Recall record Z-0372-2017 for Schaerer Medical USA Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0372-2017
Company in FDA record
Schaerer Medical USA Inc
FDA record creation date
Nov 8, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The floor locking cylinders are contained in the base of the Schaerer 7300 model surgical tables. The cylinders were manufactured with the incorrect "O" ring which may cause the cylinder to leak.

Source facts

Action
Schaerer Medical sent a Field action letters dated September 27, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. The customer is required to complete and return a response form with the number of affected tables and cylinders. All affected cylinders will be replaced in the field and returned to the manufacturer. For questions Quality Assurance at 513-272 -9116.
Initiated
2016-09-27
Root cause
Nonconforming Material/Component
Terminated
2020-07-17
Distribution
Worldwide Distribution - US Nationwide in the states of: CA, CT, FL, IA, IN, MI, MN, MO, OH, PA, and VA Replcement cylinders wre distributed
Recall number
Z-0372-2017
Recall status
Terminated
CDRH record ID
149954
FDA record created
2016-11-08
Recall event number
75288

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗