Z-0372-2020: Infinity M300 telemetry monitoring device Software versions VG2.3.1 and lower, Model Numbers MS25755, MS2

Recall record Z-0372-2020 for Draegar Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0372-2020
Company in FDA record
Draegar Medical Systems, Inc.
FDA record creation date
Nov 14, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The devices have potential cybersecurity vulnerabilities, which can include Distributed Denial of Service (DDoS), Spoofing, and Tampering.

Source facts

Action
An Urgent Medical Device Recall notification letter dated 10/9/19 was sent to customers. Recommended cybersecurity precautions: While we are in the process of updating the software, we recommend that you follow best recommended practices to limit access to the devices and wireless network: " Physical security of the patient monitors is recommended and is the responsibility of the operating organization. " Physical security boundaries to prevent access to your facility's network infrastructure is recommended and is the responsibility of the operating organization. " Drager relies on the medical device isolation mechanism of the VLANs and the proper configuration, implementation, and use of the operating organization's security measures to prevent the introduction of malware onto the Infinity Network. Action required: Please complete the attached Acknowledgement and Response Form and return it to Drager per instructions included on the form. The software upgrade will be provided free of charge. The upgrade path for your M300 device is dependent on the part number of your M300s currently in use. Your local Drager representative will contact you prior to the availability of software version VG2.3.2 to discuss the upgrade path. If you have any questions regarding this letter, please contact Michael Kelhart between the hours of 8:00 AM - 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349).
Initiated
2019-10-09
Root cause
Software design
Distribution
US nationwide distribution.
Recall number
Z-0372-2020
Recall status
Open, Classified
CDRH record ID
177026
FDA record created
2019-11-14
Recall event number
84028

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗