Skip to content

Public FDA record

Z-0372-2023: NextAR Shoulder Surgical Technique, Shoulder NextAR RSA Platform, CT based computer-assisted surgical nav

Recall record Z-0372-2023 for Medacta Usa Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0372-2023
Company in FDA record
Medacta Usa Inc
FDA classification date
Dec 3, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A warning has been added to the Surgical Technique for NextAR Shoulder, advising verification of suitable patient conditions prior to placement of the NextAR Shoulder Target Holder: WARNING Target fixation requires adequate coracoid bone stability. Osteoporosis, weakness of the coracoid process, or other deficiencies, could lead to adverse events including bone fracture. Verify that the coracoid bone and relevant patient conditions are suitable before proceeding.

Source facts

Action
Medacta issued an URGENT - Field Safety Notice to the surgeons on 10/28/2022 by email The notice provided the following information: "1. Description of advisory: The following Warning has been added to the Surgical Technique for NextAR Shoulder, advising verification of suitable patient conditions prior to placement of the NextAR Shoulder Target Holder: WARNING Target fixation requires adequate coracoid bone stability. Osteoporosis, weakness of the coracoid process, or other deficiencies, could lead to adverse events including bone fracture. Verify that the coracoid bone and relevant patient conditions are suitable before proceeding. The updated Surgical Technique is attached to this document and can be found at: resources.medacta.com. 2. Action to take: - Transmit this communication all end-users, ensuring it is fully understood. - Verify acknowledgment through end-user signature on the attached receipt. - Return all the signed receipts to the Regulatory Department of Medacta International. 3. Contact person and return address: Stefano Baj, Regulatory Affairs Director Medacta International SA, Strada Regina, CH-6847 Castel San Pietro, Switzerland Tel: +41 (0)91 696 60 60, Fax: +41 (0)91 696 60 66"
Initiated
2022-10-27
Root cause
Other
Terminated
2023-10-23
Distribution
AL, FL, NH, CO, IL, SC, AR
Recall number
Z-0372-2023
Recall status
Terminated
CDRH record ID
196706
FDA classified
2022-12-03
Recall event number
91091

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗